Skip to main content
Fig. 4 | Virology Journal

Fig. 4

From: In-house ELISA protocols for capsid p24 detection of diverse HIV isolates

Fig. 4

Comparative reactivity towards HIV-1 isolates between assays as determined by Pearson’s r correlation. ANG-, SB-, and RND-CA-p24 ELISAs exhibit a significantly high correlation with ABR-CA-p24 ELISA. ANG-CA-p24 ELISA exhibits a significantly high correlation with RND-CA-p24 ELISA, whereas both RND- and ANG-CA-p24 ELISA show a low correlation with SB-CA-p24 ELISA. Concentrations of CA-p24 are displayed in ng/mL. ABR = Aalto Bio Reagents; ANG = Anogen; SB = Sino Biological; RND = R&D Systems

Back to article page